  • Model NO.: SP10-16
Product Description
Customer type: Hospital
Equipment name: GE SP10-16 high-frequency probe
Maintenance time: 2010-4-2
Service period: 30 days
Problem Description: Image fuzzy.
Test results: Array of failure.
Solution: Replace lens, crystal, housing and strain relief.